Study uses ultrasound to improve cryo-electron microscopy samples
Researchers at École polytechnique fédérale de Lausanne (EPFL), in Switzerland, apply ultrasound during vitrification, before imaging. The result improves particle sampling without establishing a universal solution.
By Wellinton OliveiraCEO of Lyntev
Published
4 min read

Key points
- The peer-reviewed study was published in Nature Methods on 26 November.
- The comparison includes the 50S ribosomal subunit, C-reactive protein and haemagglutinin.
- A plunge-freezing experiment produced a haemagglutinin reconstruction at 2.0 Å.
- The proposed mechanism and broader application remain subjects for investigation.
Researchers have shown that applying ultrasound while preparing samples can reduce a recurring cryo-electron microscopy problem: particles becoming trapped in nearly the same orientation. The study published in Nature Methods on 26 November compares preparations with and without ultrasonic excitation in three biological systems and presents an experimental adaptation for plunge-freezing equipment.
The result concerns determining molecular structures, rather than a medical treatment. Improvements were observed in the samples and conditions tested. Applying the method to many other proteins remains the authors’ prospect, distinct from general validation or a commercial device ready for routine use.
Why particle orientation limits an image
In cryo-electron microscopy, a sample is rapidly cooled into non-crystalline ice, a process called vitrification. A microscope records images of many particles. Reconstructing a three-dimensional structure benefits from observing them from different directions.
Problems arise when particles adhere to the air–water interface through regions of their surfaces. Repeated orientations leave other views underrepresented. A reconstruction can resolve some directions well while missing information in others. A single resolution number therefore does not describe the result’s entire quality.
The team studied the 50S ribosomal subunit, a molecular complex, the pentameric form of human C-reactive protein and haemagglutinin. The latter is a particularly difficult example of preferred orientation, according to the studies discussed by the authors.
Ultrasound during vitrification, before imaging
The experimental device combines a jet-vitrification module with commercial plunge-freezing equipment. A small amount of solution is applied to a perforated grid; excess liquid is removed, then a liquid ethane jet vitrifies the sample.
A transducer positioned nine millimetres from the sample produces ultrasound at 490 kilohertz. Excitation begins approximately ten milliseconds before the jet arrives. The intervention acts during sample preparation, rather than during subsequent microscope imaging.
The proposed explanation is that support oscillations move the liquid film, helping particles detach from the interface and change orientation. The authors present tomographic observations consistent with that interpretation. They nevertheless acknowledge that some particles might rotate while remaining attached and that the dominant mechanism could depend on the protein.
What improved in each comparison
The paper uses the sampling compensation factor, SCF*, to describe how uniformly viewing directions are represented. It runs from zero to one, with one representing all directions equally. The comparisons below concern reconstructions described in the paper, rather than a clinical or industrial trial.
| Sample and measure | Without ultrasound | With ultrasound |
|---|---|---|
| 50S subunit: SCF* | 0.66 | 0.83 |
| 50S subunit: resolution | 2.8 Å | 2.6 Å |
| C-reactive protein: SCF* | 0.80 | 0.90 |
| C-reactive protein: resolution | 4.3 Å | 3.7 Å |
For haemagglutinin, preparation without ultrasound produced an incomplete reconstruction. With excitation, the team obtained a complete map at a nominal resolution of 2.7 Å. The incomplete reconstruction also had a nominal 2.7 Å resolution, illustrating why viewing directions and the material actually reconstructed matter.
An analytical caveat is also important. Particle selection during processing favours particular views. When estimating orientations across broader sets, the authors obtain SCF* values of 0.28 without ultrasound and 0.50 with it. These must not be confused with higher measures for particles selected for the final reconstruction.
A plunge-freezing adaptation reached 2 Å
The original supplement describes a proof-of-concept implementation for plunge freezing. The grid passes several transducers as it descends. In that haemagglutinin experiment, the researchers report a 2.0 Å reconstruction and a more favourable orientation distribution.
An ångström is one tenth of a nanometre. The stated resolution describes the reconstructed structure and its processing, rather than the protein’s overall size. The experiment demonstrates a route towards adapting existing devices, without establishing that every such device can accommodate the modification without further development.
Other proteins and validation beyond this device remain necessary
The work is a peer-reviewed scientific article, accepted on 27 October and published on 26 November. Its methods and supplement detail preparation, image acquisition and computational selection. The 50S tomographic comparison analyses three tomograms per condition when determining particle distribution, defining the scope of that set of observations.
Generalisability to other systems, operation in other laboratories and optimum excitation conditions remain to be demonstrated. The authors propose tuning frequency and power further. The current result is an experimental improvement in preparation and reconstruction, with identified limits, rather than a promise to solve every difficult sample.
Five authors disclose filing a patent application in Europe covering sample preparation on 7 August 2025. They also state that Ulrich J. Lorenz will be part of Alpion CryoSolutions, a company being formed that plans to commercialise the technology. The disclosure identifies interests in its development; it does not establish that a patent has been granted or a product launched.
Sources
Documents and reporting consulted for this story.
- Mitigating preferred orientation in cryo-EM with ultrasonic excitation during vitrificationNature Methods — Harry M. Williams e coautores, . Primary source. accessed 27 November 2026
- Supplementary Information — Notes1–6 and tablesHarry M. Williams e coautores. Primary source. accessed 27 November 2026