Wrist-sensor study exposes limits in cardiac-arrest detection

The model detected 48.5% of events without false alarms in the daily-life recordings analysed. These are preliminary data without emergency-response or survival testing.

By CEO of Lyntev

Published
3 min read

A black Max-Health-Band wristband on a wrist in a photograph taken in 2020.
Max-Health-Band for acquiring optical wrist PPG signals, photographed in 2020. It illustrates the measurement principle and location; it is a different device, not the study’s Philips Elan wristband or BECA system. Peter H. Charlton, CC BY 4.0; full image resized. Peter H. Charlton / Wikimedia Commons (opens in a new tab). CC BY 4.0; redimensionamento integral, sem recorte. (opens in a new tab)

Key points

  • The 26 November manuscript has not undergone peer review.
  • Analysis covered 33 hospital-induced events and 100 healthy volunteers.
  • Higher sensitivity came with more false alarms.
  • The study separates procedural artefacts from spontaneous loss of circulation.

A preliminary Dutch study tested a model that seeks to recognise loss of pulse from optical wrist sensors. Under the conditions evaluated, it detected 48.5% of hospital episodes without producing false alarms in the daily-activity recordings used for comparison. This does not demonstrate that a watch can safely alert the general population to cardiac arrest.

The BECA-S2/BECA-S6 manuscript was posted on medRxiv on 26 November and has not undergone peer review. The team combined signals from 31 patients undergoing cardiac procedures with recordings from 100 healthy volunteers. The experiment also exposes a methodological risk: an algorithm may learn traces of the hospital procedure rather than recognise loss of circulation alone.

An optical signal in two different settings

Photoplethysmography, or PPG, uses light to track changes in blood volume beneath the skin. Wrist devices use this principle, but the original study (opens in a new tab) collected data with Philips Elan research wristbands. It did not evaluate an application available for installation on a consumer watch.

The wristbands recorded green-light PPG and movement through accelerometers. Hospital participants underwent ventricular-arrhythmia ablation or testing during subcutaneous defibrillator implantation. Loss-of-pulse episodes were induced as part of those procedures and monitored with an electrocardiogram and, where available, arterial blood pressure.

Two specialists annotated events in the clinical signals and resolved disagreements by consensus. The analysis population comprised 31 patients; 22 experienced a total of 33 episodes. Exclusions included episodes lasting under three seconds. These were controlled hospital situations rather than spontaneous cardiac arrests at home or in the street.

To assess false alarms, the team used a separate group of 100 healthy volunteers performing unsupervised daily activities for approximately one hour while wearing a wristband on each wrist. Usable recording totalled 156.7 wrist-hours. This unit counts both wrists; it does not represent following 100 people for several days.

Detecting more events brought more alarms

The model was assessed with cross-validation, separating participants between training, parameter selection and testing. Each of 22 splits held out one patient with events for testing. Healthy volunteers were also distributed across the splits. This was therefore not an independent prospective validation in a new population after the algorithm had been fixed.

At the operating point with no observed false alarms in daily-life recordings, sensitivity for hospital events was 48.5%. Its 95% confidence interval ranged from 32.1% to 65.5%, reflecting uncertainty in a small sample. No false alarms during the observed period does not establish a zero rate over prolonged use.

Adjusting the threshold to recognise more episodes increased false alarms. At 80% sensitivity, the estimated rate was 7.36 false alarms per hour. The 95% sensitivity target was not reached within the evaluated range, which extended to 100 false alarms per hour.

The findings show why a single classification percentage is insufficient to assess a potential emergency system. The balance between missed events and unnecessary alerts has practical consequences. This work did not test automatic calls, emergency-response times or improved survival.

When a model learns the wrong clue

Electrical arrhythmia induction and defibrillation shocks leave traces in the signals. Those traces would not necessarily occur during a spontaneous cardiac arrest. A model associating them with an event could appear more effective in hospital than it would be elsewhere.

The authors masked segments containing artefacts and added a training penalty to reduce dependence on the masked intervals. Without this mitigation, artificial clues shifted model scores by an amount corresponding to 35.3% of their discriminative range. This is not a 35.3-percentage-point improvement in a device’s accuracy.

What remains to be established outside hospital

The team calls for more representative data from spontaneous arrests and longer observation. Most patients had light skin, and the groups differed in clinical and demographic characteristics. The analysis does not establish equal performance across all skin tones or conditions.

The contextual photograph shows a Max-Health-Band, another real device acquiring wrist PPG, photographed in 2020. It illustrates the measurement location and principle without depicting the Philips wristband, BECA algorithm or a clinical alarm.

Funding came from a public–private partnership through the Dutch Heart Foundation and Philips; one co-author is a Philips employee. The authors provide research code (opens in a new tab), while clinical data access depends on consent and sharing conditions. Lyntev News did not run the model or reanalyse the recordings. The publication adds preliminary evidence and validation limits, rather than a recommendation for clinical use.

Sources

Documents and reporting consulted for this story.

  1. Smartwatch-based loss-of-pulse detection for out-of-hospital cardiac arrest: the BECA-S2 and BECA-S6 studies (opens in a new tab)Roelof G. Hup e colaboradores / medRxiv, . Primary source. Accessed 27 November 2026
  2. Cardiac arrest detection from wrist PPG and accelerometry (opens in a new tab)Roelof G. Hup / GitHub. Primary source. Accessed 27 November 2026
  3. Max Health Band — fotografia própria do autor (opens in a new tab)Peter H. Charlton / Wikimedia Commons, . Primary source. Accessed 27 November 2026

Editorial principlesReport an error

Topics

Wellinton Oliveira

Written by

CEO of Lyntev

Technology with context, original analysis and sources you can consult.